13 Calibrating Redox sensors
32
13.3.2
Two-point Calibration
Step
Action
1
Ensure that
•
the calibration default settings are correct
•
"Percent" is set as the Redox unit in the configuration of the Re-
dox measuring input.
v
Chapter 13.2.2 "Calibration default settings for Redox sensors",
v
Chapter 9.5.3 "pH/Redox/NH3 analysis inputs", Page 102
2
Start the two-point calibration.
Device Menu
r
Calibration
r
Select Analysis Input for pH/Redox/
NH
3
r
Two-point calibration
3
Enter the concentration value of the first reference solution in per-
cent. Confirm by pressing "OK".
4
Clean the Redox electrode and immerse it in the first test solution.
Wait until the measuring value displayed stabilizes and then con-
firm the result of the measurement by pressing "OK".
5
As in step 3, enter the concentration value of the second solution
in percent. Confirm by pressing "OK".
6
Clean the Redox electrode and immerse it in the second
test solution. Wait until the measuring value displayed stabilizes
and then confirm the result of the measurement by pressing "OK".
7
A protocol summarizing the calibration values determined then ap-
pears. Acknowledge the protocol by pressing "OK".
Failed calibrations are canceled at this point and discarded.
8
Press "Yes" to accept the calibration values determined and enter
the calibration in the calibration logbook.
Press "No" to discard the results.
Tap button to enter
field
Summary of Contents for 202580
Page 6: ...Content...
Page 28: ...12 Calibrating a pH measuring chain 28...
Page 42: ...15 Calibrating CR conductivity sensors 42...
Page 52: ...16 Calibrating Ci conductivity sensors 52...
Page 60: ...17 Calibrating universal inputs 60...
Page 79: ...79 18 Technical data...
Page 80: ...18 Technical data 80...
Page 81: ......