"Connect unit" menu
66
2.13.2018
10.5.2. JULABO Sensor Pressure / Flow
Attention
The socket may be used with original JULABO accessories only.
Any other use may damage the unit's electronics.
External sensor signal
Switch the external sensor signal
on and off.
The
value
to select the sensor
connected to the JULABO
Pressure/Flow Sensor jack.
>
Off
<,
>
Pressure
<, signal from an
external pressure sensor, or
>
Flow
< signal from an external
flow sensor.
The
filter time
is set to 1 second
as default. The filter time may be
increased (0 - 1000 s) to display
the value with greater stability.
Note:
This will increase the
reaction time if the system is
controlled by the pressure or flow
value.
Sets the measurement which
corresponds to a
4 mA input
signal
Sets the measurement which
corresponds to a
20 mA input
signal
The value on the left is the current measurement at the present
time whilst the one on the right is the pressure or flow value
calculated using the current settings.
Preparations
Connect the external sensor to the 4-pin jack on the rear of the
device.
To display the value, refer to page 20