"Determine thermodynamics" menu
32
2.13.2018
5.2.1.3.
Type „Flow rate Control“
Select
Set value.
Note
Check the selected pressure limits!
(Page 48)
The selected pressure limits are
monitored during flow control as
well.
A high flow rate may exceed these
limits and cause the unit to shut
down.
The flow rate is infinitely adjustable and is actively controlled.
The setting range depend on the performance category of the
temperature control system.
Example: Setting range see display of unit.
Example: 12.00 l/min
Refer to the operating manual of the utilized flow control unit
for additional notes on possible flow rates.
5.2.2. "Pump Mode" menu
Select
>
Pump Auto
<
The pump is controlled via the
start/stop button or via the
interface.
>
Pump on
<
Pump runs continuously.
>
Pump after-run
<
You must select the pump’s after-
running time.
>
Pump after-run
<
Set time.
Example: 5 minutes
hh:mm:ss