![Julabo Magio MX Original Operating Manual Download Page 99](http://html1.mh-extra.com/html/julabo/magio-mx/magio-mx_original-operating-manual_2030737099.webp)
Appendix
99
40102
101
ushort
Month
40103
102
ushort
Day
40104
103
ushort
Hour
40105
104
ushort
Minutes
40106
105
ushort
Second
11.3.5 Input register
Register
address
Protocol
address
Data type Explanation
Value range
30001-
30002
0-1
uint
Firmware version
Byte1: Major
Byte2: Minor
Byte3: Build
Byte4: Revision
30004-
30005
3-4
uint
Barcode
30006
5
short
Remote control mode
30007
6
short
Alarm code
30008
7
short
Warning code
30011-
30012
10-11
float
Bath temperature
30013-
30014
12-13
float
Temperature value of the
external Pt100 temperature
sensor
30015
14
short
Present heating
capacity (%)
-100 … 100
30016-
30017
15-16
float
Safety sensor temperature
value
30042
41
short
Current bath fluid filling
level (%)
0 ... 100
30051-
30052
50-51
float
Si - Gradient of the internal
bath
30053
52
ushort
Ti - Time constant of the
internal bath
30054
53
ushort
Te - Time constant of the
external bath
Summary of Contents for Magio MX