![Julabo Magio MX Original Operating Manual Download Page 43](http://html1.mh-extra.com/html/julabo/magio-mx/magio-mx_original-operating-manual_2030737043.webp)
Initial operation
43
NOTE
Overflowing bath fluid due to externally connected systems!
If the externally connected system is higher than the temperature control
system, bath fluid can flow back and overflow when switched off.
• Position the connected external system on the same level or lower than
the temperature control system
• Position a shut-off valve or Solenoid valve between the external system
and the temperature control system as backflow protection.
NOTE
Damaged hoses due to kinking!
Hoses are damaged by kinking and can may leak.
• Lay hoses with large radii
• Avoid kinking of the hoses
7.3.1 Connect an external system with screw connections
This section describes how to connect an external, closed system to the device
using screw connectors.
Open-end wrench, 19 mm
The temperature control hoses of the external system are equipped with
M16x1 female and barrel gaskets.
1. Disassemble the union nuts on the pump connectors.
2. Remove the sealing plugs.
Summary of Contents for Magio MX