![Julabo Magio MX Original Operating Manual Download Page 40](http://html1.mh-extra.com/html/julabo/magio-mx/magio-mx_original-operating-manual_2030737040.webp)
Initial operation
40
7.1.2 Mount heating or refrigerated circulator
In the case of a heating circulator or a refrigerated circulator, the circulator is
mounted on the closed bath or on cooling machine.
If the circulator is disassembled, e.g. due to a change of device or for service
purposes, it can then be easily reassembled with the connecting box.
Open-end spanner, size 7 mm
The thermostat is ready to be mounted on a closed bath or a cooling
machine.
The bath is empty.
1. Carefully insert the circulator into the bath.
Align the four threaded sleeves of the bath with the four fastening screws of
the connecting box.
2. Tighten the four fastening screws [arrows in the image right] by hand.
3. Tighten the fastening screws evenly with the open-ended wrench.
4. Check that the circulator is sitting correctly and that the connecting box
gasket is seated gap-free on the surface of the unit.
The heating circulator or refrigerated circulator is mounted.