Product description
19
5.4.2 Stakei output
The Stakei output is a control output that can be adjusted to solenoid valves
with various tasks via the operating menu. The desired function is set and
activated in the operating menu.
The Stakei output can be assigned the following functions:
• Refill device:
The Stakei output switches on when a low liquid level warning is triggered.
If the level of the bath fluid drops below a certain level, the Stakei output
turns on and a connected refilling device can be activated. This pumps bath
fluid into the bath tank until the filling level has returned to a non-critical
level.
• Cooling pulse:
This function is controlled by the current cooling capacity of the device. If
the temperature setpoint is lower than the actual temperature detected by
the device, a pulse-width modulated signal is sent via the Stakei output.
This signal can be used to control a solenoid valve for cyclic counter-cooling
via a cooling coil.
• Backflow:
This function is used as backflow protection in a system where the
connected application is physically higher than the circulator. When the
pump is running, the Stakei output is switched on and a check valve
inserted between the circulator and the application can be controlled. This
prevents the bath fluid flowing from the application to the bath tank while
the pump is not running and causing the bath tank to overflow.
Technical data
Stakei output
Mains connection 100-115 V
100-115 V/1 A
Mains connection 200-230 V
200-230 V/1 A
Summary of Contents for Magio MX