Operation
46
8 Operation
8.1 Switch on the unit
This section describes how to switch on the device.
v
The device is connected and ready for operation.
1. Switch the device on at the mains switch.
All display elements light up briefly, the software boots and starts the
device.
The device is switched on and ready for operation. It switches to the most
recently activated operating mode, manual operating mode or remote
control mode. If the auto start function is activated, then the device starts
directly into the last setting.
8.2 Switch off the unit
This section describes how to switch off the device.
The device is switched on.
1. Stop a running temperature control application.
If a cooling machine is connected, the cooling symbol flashes and the
cooling machine is shut down.
Do not switch off the device until the cooling symbol stops flashing and the
device is in standby mode.
2. Switch the device off at the mains switch.
The device is switched off.
8.3 Configuring setpoint temperature
Device is running the temperature control application to the configured setpoint
temperature. The factory setting is 10°C. The setpoint temperature can be
changed while the temperature control application is running. The set value is
saved.
The device is switched on.
1. Press one of the arrow keys briefly.
The display switches from the actual value display to the setpoint display.
The digits before the decimal point flash.
2. Use the arrow keys to set the value before the decimal point and confirm
with
[OK]
.
The set value is applied. The decimal point flashes.