Operation
49
The set value is applied. The decimal point flashes.
3. Use the arrow keys to set the value after the decimal point and confirm with
[OK]
.
The set value is applied. The new setpoint temperature flashes briefly.
The setpoint temperature is set and active. Setpoint temperature
configuration can be interrupted using the
[MENU]
key. The previous value
is then retained.
8.4 Start temperature control application
A temperature application can be started directly on the unit. Other options
include timer-controlled temperature control application with the integrated
programmer and remote control via a connected PC.
The unit is ready for use.
1. Switch the unit on at the mains switch.
2. Use the arrow keys to set the desired setpoint temperature.
3. Press and hold the
[OK]
key until the temperature control application starts.
The setpoint temperature is saved. The display flashes briefly. The unit starts
the temperature control application at once. The temperature control
application can be stopped with the
[OK]
key.
Observe the following for heating circulators:
For temperature control applications near or below the ambient
temperature: Use a cooling coil or JULABO immersion cooler.
8.5 Activate autostart function
The autostart function makes it possible to start a temperature control
application directly using the mains switch or via an intermediate timer.
The device is configured ex works in such a ways that it switches to a safe
operating status in the event of power failure. The autostart function is
deactivated. The display shows “OFF.” The refrigeration aggregate, heater, and
pump motor are disconnected from the mains voltage.
The device is switched on.
1. Press the
[MENU]
key.
2. Use the arrow keys to select the menu option
<conF>
and confirm with
[OK]
.
3. Use the arrow keys to select the menu option
<Auto>
and confirm with
[OK]
.