Initial operation
39
7.3 Connecting an external system
The device is designed for tempering external, closed loop systems. An external
system is connected to the unit’s pump connections.
CAUTION
Risk of burns due to damaged temperature control hoses!
Hot bath fluid can escape from damaged temperature control hoses and
cause serious burns when it comes into contact with skin.
• Check the temperature control hoses regularly for integrity
• Immediately replace damaged temperature control hoses
• Do not kink temperature control hoses
• Regularly exchange temperature control hoses
• Check the pump connections for leak tightness
NOTE
Material damage due to incompatible externally connected
system!
If the temperature range and/or pressure parameters of an externally
connected system do not match those of the device this can result in damage
to individual components or even failure of the entire system.
• Before connection check the external system for compatibility with the
device combination
• If an external system is connected that is not set up for the maximum
pressure of the device, the pump's flow rate must be limited in the
settings
• If an external system is connected, the safety of the entire system is the
responsibility of the operator
NOTE
Hot pump connections!
The pump connections can become very hot during operation. Heat-sensitive
parts or cables can be damaged if they have contact.
• Pump connections must be uncovered during operation
• No loose parts or cables should come into contact with pump connections
during operation