©JRI Maxant
22
a
a) Unplug the battery connector
.
b) Short circuit the 2 pins with a screwdriver
c) Connect the new battery
d) Change the seal
e) Close the battery cover
f) Apply configuration using Sirius Software to
initialise the unit
Never unscrew. This
will cancel the
warranty
1
3
Open the
battery cover
b
1
2
3
2
Device set up in transportation mode
XVII.
BATTERY CHANGE
When the SPY RF battery has to be replaced, the LCD screen displays the following message:
To replace the battery, follow the instructions below:
XVIII.
RESET
If the device does not work anymore (cannot turn it on…), use the Reset function in the same way as the
battery change.
XIX.
MAINTENANCE
Clean with a soft cloth, dry or lightly moistened with water. To remove stubborn dirt, use a soft cloth
impregnated with a mild detergent, non-abrasive. Then wipe gently with a soft, dry cloth.Never use benzene,
thinners, alcohol or solvents of any kind, which may cause discoloration or deformation of surfaces
Pushbutton pressing
Mode
< 5"
5"<appui>10"
Off
-
The 2 leds are on and flash at
the same time
.
Starting measurements
Pushbutton
Green led 2" = beginning of
measurements
-
Delayed (date & time)
The 2 leds are on and flash at
the same time
= Waiting for
starting measurements
Immediately
-
The 2 LEDs are on and then
flash alternatively
= ending
measurements
Mesure
Green led 10" = Top zone
The 2 LEDs are on and then
flash alternatively
= ending
measurements
DOWNLOAD THE MEMORY BEFORE CHANGING THE BATTERY.
NEVER WAIT UNTIL THE BATTERY IS EMPTY OR THE DATA
WILL BE DELETED.
KEEP THE BATERY FAR FROM THE FIRE ; DO NOT TRY TO RELOAD OR TO SHORT CIRCUIT IT. USE ONLY BATTERIES
SUPPLIED BY JRI (REF 06569)