Jorvet Vet Pro Pump Maintenance & Troubleshooting Download Page 1

 
1.   Turn pump 

ON 

2.   Press 

INFO

 key twice  

3. Scroll down with the down- 

arrow key 

(0

)

  

       to 

ADJUST SETTINGS

.

 

4.   Press 

START/OK

 key. 

5. Scroll down with the down- 

arrow key 

(0

      to 

MORE

.  

 

6.   Press 

START/OK

7. Using the numeric key pad, input Technician 

Code 

901 

8. Press 

START/OK

 key. 

9. Scroll down with the down- 

arrow key 

(0

)

  

       to 

MAXIMAL RATE

.  

10. 

Press 

START/OK

11. 

I

nput 

1200

 ml. 

12. Press 

START/OK

 key. 

13. Scroll down with 

arrow 

       keys (2

, 0

) to   

      

RESTART PUMP

.  Confirm by  

      pressing  

START /OK

.

 

Issue: 

Pump will not rate entry above 500 ml/ hr  

A small numeral 3 will be displayed on the upper  
right-hand corner of the LCD Display when the Factory  
Defaults have been changed. 

Solution:  

 Reset factory default Maximal Rate. 

The Vet 

Pro Pump

 

comes with a

 factory default Maximal Rate of 

500ml. 

 

Maintenance & Troubleshooting 