1. Turn pump
ON
2. Press
INFO
key twice
3. Scroll down with the down-
arrow key
(0
)
to
ADJUST SETTINGS
.
4. Press
START/OK
key.
5. Scroll down with the down-
arrow key
(0
)
to
MORE
.
6. Press
START/OK
.
7. Using the numeric key pad, input Technician
Code
901
8. Press
START/OK
key.
9. Scroll down with the down-
arrow key
(0
)
to
MAXIMAL RATE
.
10.
Press
START/OK
.
11.
I
nput
1200
ml.
12. Press
START/OK
key.
13. Scroll down with
arrow
keys (2
, 0
) to
RESTART PUMP
. Confirm by
pressing
START /OK
.
Issue:
Pump will not rate entry above 500 ml/ hr
A small numeral 3 will be displayed on the upper
right-hand corner of the LCD Display when the Factory
Defaults have been changed.
Solution:
Reset factory default Maximal Rate.
The Vet
Pro Pump
comes with a
factory default Maximal Rate of
500ml.
Maintenance & Troubleshooting