Joerns Lifting and Repositioning
HPL402 Hoyer Power Patient Lift

© 2015 Joerns Healthcare • 000-3558 RevD • 15-2003

6

Hoyer 4-point Sling Information

Warning:

 Torn, cut, frayed or broken slings could 

fail resulting in serious injury or death to patient. 
Use slings in good condition only. Destroy and 
discard old, unusable slings.

Note:

 Do not use any other manufacturer’s sling 

with the HPL402 Lifter.

Note:

 Do not use Hoyer slings with any other 

manufacturer’s lift.

Patient Size Guide

A

B

Weight

inch (cm) inch (cm) lbs (kg)

Large

28 (70)

19 (47)

198-400 (90-
191)

Medium 24 (61)

19 (47)

99-210 (45-95)

Small

18 (46)

15 (38)

55-110 (25-50)

*Patient’s weight to sling size information is an 
approximation only.

Patient can be raised 4" higher by hooking up the 
inside strap to the cradle.

Example of label found on back of Partner 
Slings

Summary of Contents for HPL402 Hoyer

Page 1: ...User Manual Joerns Lifting and Repositioning HPL402 Hoyer Power Patient Lift To avoid injury read user s manual before using ...

Page 2: ... center may wear prematurely Warn cradle parts could fail resulting in injury to resident caregiver Hanger post should protrude cradle center by 1 16th inch Figure 1 1 Carefully unpack lifter in the following order a Remove the outside top of the box b Remove the three inside top fillers from the box NOTE Be careful with the center filler it contains the electronics c Remove base and adjustment ha...

Page 3: ...er Post is flush or recessed excessive wear may have occurred Worn cradle parts could fail resulting in serious injury or death to patient and or caregiver Remove from service and replace worn parts immediately Handset Battery Pack Wire Clips 2 Opposite Side Lock Nut 4 Boom Mast Lock Bolt Shoulder Bolt 4 Power Cord Actuator Washer 4 Caster Wheel Hanger Post Thrust Washer Mast Screw Cntrol Box Mtg ...

Page 4: ...ustic battery alarm at about 50 battery capacity 0 LCD Display Troubleshooting Power Unit Symptom Possible Cause Action Power indicator does not light up Not connected to mains The fuse has blown Control box defective Connect to mains Replace fuse if the system is prepared for external fuse replacement or send the system for repair Send control box for repair Power indicator lights up but actuator...

Page 5: ...he control box must always be noted If the cycle has been exceeded the control box could overheat and become damaged Unless otherwise specified on the label the duty cycle is maximum 10 or six 6 min hour of continuous operation Emergency stop button should be pushed in when not in use for an extended period of time the LCD Display will trickle down the battery Note Power lifter will not be operabl...

Page 6: ...rd old unusable slings Note Do not use any other manufacturer s sling with the HPL402 Lifter Note Do not use Hoyer slings with any other manufacturer s lift Patient Size Guide A B Weight inch cm inch cm lbs kg Large 28 70 19 47 198 400 90 191 Medium 24 61 19 47 99 210 45 95 Small 18 46 15 38 55 110 25 50 Patient s weight to sling size information is an approximation only Patient can be raised 4 hi...

Page 7: ...ked into mast sleeve Inspect for bends or deflections Inspect all bolts and nuts for wear and tightness Boom Inspect all bolts and nuts for wear and tightness Inspect to ensure the boom is centered between the base legs Cradle Check hanger post for wear or damage Replace post if wear exceeds 1 16 Check sling attachments for wear or deflection Slings and Hardware Check metal parts for wear or damag...

Page 8: ...structions is not covered Any alteration modification or repair unless performed by or authorized in writing by Joerns Healthcare will void this warranty This warranty applies in the USA only Check with your supplier to find out if international warranties apply Parts Hoyer Classics Lifts contain various parts that wear from normal use These parts such as rubber parts plastic parts and painted sur...

Reviews: