JK2830 Operation Manual Chapter 1 Out of Box Audit
2
1.4
Environment
1)
Please do not operate the instrument in the place that is vibrative, dusty,
under direct sunlight or where there is corrosive air.
2)
The normal working temperature is 0
℃
~
40
℃
, relative humidity ≤75%, so
the instrument should be used under above condition to guarantee the
accuracy.
3)
There is heat abstractor on the rear panel to avoid the inner temperature
rising. In order to
keep good airiness, please don’t obstruct the left and right
airiness holes to make the instrument maintain the accuracy.
4)
Although the instrument has been specially designed for reducing the noise
caused by ac power, a place with low noise is still recommended. If this
cannot be arranged, please make sure to use power filter for the instrument.
5)
Please store the instrument in the place where temperature is between 5
℃
and 40
℃
, humidity is less than 85%RH. If the instrument will not be put in
use for a time, please have it properly packed with its original box or a similar
box for storing.
6)
The instrument, especially the test cable should be far from strong
electro-magnetic field, to avoid the jamming on measurement.
1.5
Use of Test Fixture
Please use the accessory test fixture or cable,
the test fixture made by user or
from other company may cause the incorrect measurement result
. The
test fixture or cable should be kept clean, as well as the pin of DUT, thus to
guarantee the good connection between DUT and fixture.
Connect the fixture or cable to four test terminals Hcur, Hpot, Lcur, Lpot on the
front panel. As for the DUT with shielding shell, connect shielding layer or ground
“
”.
Note: When test fixture or cable has not being installed, the instrument will
display an unstable test result.
1.6
Warm-up
1)
To guarantee the accurate measurement, the warm-up time is no less than
15min.
2)
Please not turn on or off instrument frequently, in order to avoid the inner
data fluster.