JK2830 Operation Manual Chapter 5 Execute LCR operation and some examples
6.6.9
Accuracy of Z
Test condition:
Function Z-
θ
Test frequency 100Hz 1kHz 10kHz 100kHz Test respectively
Level 1V
Range AUTO
Bias 0V
Speed Slow
Open and short correction should be made before testing. Connect standard AC
resistors: 10Ω, 100Ω, 1kΩ, 10kΩ, 100kΩ and change the freq
uency. The error
between reading and nominal value should be in the range ruled in this chapter.
65