JK2830 Series Operation Manual Chapter 3 Introduction to [DISP]
31
Execute the following steps to set sorting limits.
1)
Set
FUNC
and
NOM
in the compare function menu and the limit
MODE
of the
primary parameter.
2)
Move the cursor to Low limit of BIN 1. If you select percent tolerance mode,
the following operation steps should be from step 3 to step 6; if you select
absolute tolerance mode and sequential mode, the following operation steps
should be from step 7 to step 11.
3)
User numeric keys to input low limit value in Low limit. After inputting the
data, you can use (
m
,
k
,
*1
) to replace
[ENTER]
to input the limit value.
After inputting limit value in
LOW
of BIN 1, the low limit of BIN 1 will be
automatically set as
–
(absolute limit) and the high limit will be + (absolute
limit).
4)
The cursor will automatically move to
LOW
of BIN 2. Repeat step 3 until the
limits of BIN 3 are input. Then the cursor will automatically move
LOW
of 2
nd
.
5)
After inputting the low limit of the secondary parameter, the cursor will
automatically move to
HIGH
of 2
nd
.
6)
Input the high limit of the secondary parameter.
7)
In Low limit of BIN 1, use numeric keys to input the low limit. After inputting
the data, you can use (
p
,
n
,
µ
,
m
,
k
,
M
,
*1
) to replace
[ENTER]
to input the
limit value. When [ENTER] or
*1
is used to input the limit value, the default
unit
is F, H or Ω.
8)
After inputting the low limit of BIN 1, the cursor will automatically move to
HIGH
of BIN 1. Input the high limit of BIN 1.
9)
The cursor will automatically move
HIGH
of BIN 2. For the limit mode is
sequential mode, the low limit of BIN 2 will be the high limit of BIN 1. Input
the high limit of BIN 2.
10)
Repeat step 9 until the high limit of BIN 9 is input. Then the cursor will
automatically move to
LOW
of 2
nd
. Input the low limit of the secondary
parameter.
11)
The cursor will automatically move to
HIGH
of 2
nd
. Input the high limit of the
secondary parameter.
3.6
<CORRECTION>
Open and short correction can be used to eliminate the distribution
capacitance, spurious impedance and other measurement errors. JK2830
provides two correction modes: the first one is executing open and short
correction on the <MEAS DISP> page; the other one is executing open, short
and load correction on the <CORRECTION>page.
Press
[SET]
to select
CORRECTION
to enter into the
<CORRECTION>
page.