PDL5 User Manual
PDL5-UM-00001 Rev 001
70
JGR Device-Specific Command Description
PDL
The PDL command will switch the meter to PDL
mode.
Interface:
GPIB and RS-232/USB
Syntax:
“PDL”<CR>
Returned Format:
None
PDL?
The PDL? command will cause the meter to transmit
the last PDL value. In triggered PDL mode this
command will not initiate a new measurement cycle.
Use together with TRIG in triggered mode to obtain a
new reading.
Interface:
GPIB and RS-232/USB
Syntax:
“PDL?”<CR>
Returned Format:
“xx.xx” or “xx.xxx” depending on resolution
LAV?
The LAV? command will cause the meter to transmit
the last Average Loss value. In triggered PDL mode
this command will not initiate a new measurement
cycle. Use together with TRIG in triggered mode to
obtain a new reading.
Interface:
GPIB and RS-232/USB
Syntax:
“LAV?”<CR>
Returned Format:
“xx.xx” or “xx.xxx” depending on resolution
FP
The FP command will move the Flip-In Polarizer IN or
OUT of the optical path.
Interface:
GPIB and RS-232/USB
Syntax:
“FP x”<CR>
Where x is a numeric value [0|1] OR a string
[OFF|ON], stating whether the polarizer is in the
path or not.
Example “FP ON” moves the Flip-In Polarizer into the
optical path
Returned Format:
None