TMPM5200-2
5-21
5.3.2 Resetting the Controller
The SPM controller is initialized by signals from the software when the WinSPM
software is started. Before starting the software, it is necessary to ensure that the SPM
controller is on. If you switch on the SPM controller after starting the software, or switch
off the SPM controller then switch it on again after replacing the scanner, for example,
you must reset the system from the software.
To reset the SPM controller, carry out the following procedure:
1.
Select SPM Scan from the main menu, and select Reset SPM from the
pull-down menu.
The “Reset SPM?” window appears.
2.
Confirm that the SPM controller is switched on; then click on the OK button.
5.3.3 Confirming Settings of the Device Setup Dialog Box
Click on SPM Scan in the main menu bar to open the pull-down menu. Click
on Device Setup to open the Device Setup dialog box. Then make sure that
the models of the instrument, unit and attachments are correctly selected.
The basic settings of WinSPM vary according to the instrument selected. There may
be times when the system malfunctions, unless they are correctly selected. Likewise,
the piezo-element power supply voltage depends on the instrument—
±
150 V or
±
200 V. If you ignore the confirmation, there is the possibility of damaging the
system.
The settings of the Device Setup Dialog Box are preset in the factory. Should you
nSPM using the “Admin” password.
5.3.4 O
Prepare for measurement as follows:
1.
e Setup dialog box.
he Scan Mode Setup dialog box can also be opened either by pressing the S key
while holding down the Ctrl key on the keyboard, or by clicking on the measure-
ment icon.
change the setting, you must enter Wi
If any of the settings are incorrect, terminate the software once, correct them and
reboot the software using the “Admin” password.
pening the SPM (Measurement) Parameters Window
Click on SPM Scan in the main menu bar to open the pull-down menu. Click
on Scan to open the Scan Mod
T
(Measurement
2.
Click on the radio button that corresponds to the desired mode, and then
click on the OK button. The selected mode starts and the SPM Parameters
window opens.
not appear the next time you start WinSPM. The selected
mode automatically starts. It is therefore expedient to check the box when the
icon)
Check the box for “Don’t show this window when start up” and the Scan Mode
Setup dialog box will
mode you use is limited to a particular one.