![JDS Uniphase CellAdvisor JD780A Series User Manual Download Page 514](http://html1.mh-extra.com/html/jds-uniphase/celladvisor-jd780a-series/celladvisor-jd780a-series_user-manual_3602616514.webp)
JD780A Series Analyzers
514
Using Cable and Antenna Analyzer
3.
Press the
Data Points
soft key to change the resolution of your measurement, and then select
the data point option:
126
,
251
,
501
,
1001
, and
2001
. Changing the data point does not affect
current calibration.
NOTE
The larger number you choose, the higher resolution you get and the longer the
instrument takes to sweep and display results. Selecting the data point larger than what
you need for a measurement will result in unnecessarily long sweep time. It is
recommended that you select high resolution data points only for an instance of
measuring wide frequency bands or requiring precise measurement data.
4.
Optional
. To select the output power, toggle the
Output Power
soft key between
0 dBm
and
-30
dBm
.
NOTE
When an amplifier’s reflection is measured, it is recommended that you change this
output power level to -30 dBm and then re-calibrate. Changing to -30 dBm may result in
reduced measurement dynamic range.
Measurement example
Figure 265
2-port scalar measurement
Analyzing measurements
Setting trace and display
You can display up to six traces on the measurement chart simultaneously.
Procedure
1.
Press the
TRACE/DISPLAY
hard key.
2.
Press the
Select Trace
soft key, and then select the trace number:
T1
,
T2
,
T3
,
T4
,
T5
, or
T6
.
The legend shape of the selected trace changes from square to round to indicate that the trace is
the active one now.
3.
Toggle the
Trace View
soft key and select
On
to display the selected trace.
Summary of Contents for CellAdvisor JD780A Series
Page 52: ...JD780A Series Analyzers 52 Getting Started Figure 6 Example of Web remote on Chrome...
Page 254: ...JD780A Series Analyzers 254 Using LTE FDD Signal Analyzer 3 RS SINR 4 S SS RSSI 5 P SS Power...
Page 323: ...JD780A Series Analyzers Using LTE TDD Signal Analyzer 323 3 RS SINR 4 S SS RSSI 5 P SS Power...
Page 390: ...JD780A Series Analyzers 390 Using EV DO Signal Analyzer 3 Pilot 4 MAC 64 5 MAC 128...
Page 525: ......
Page 590: ...JD780A Series Analyzers 590 Appendix Figure 306 Append keywords window...