![JDS Uniphase CellAdvisor JD780A Series User Manual Download Page 512](http://html1.mh-extra.com/html/jds-uniphase/celladvisor-jd780a-series/celladvisor-jd780a-series_user-manual_3602616512.webp)
JD780A Series Analyzers
512
Using Cable and Antenna Analyzer
Measurement example
Figure 263
Smith chart measurement
Performing 2-port vector measurements
The Vector measurement is used to perform faster and more accurate measurement with around 80
dB dynamic range. It also provides antenna isolation measurement. You can determine DUT
’s S
21
phase characteristic by selecting the measurement display type as Phase.
Making a measurement
Procedure
1.
Make a
proper cable connection as described in “Connecting a cable for 2-port vector
2.
Press the
MEASURE SETUP
hot key.
3.
Press the
Data Points
soft key to change the resolution of your measurement, and then select
the data point option:
126
,
251
,
501
,
1001
, and
2001
. Changing the data point does not affect
current calibration.
NOTE
The larger number you choose, the higher resolution you get and the longer the
instrument takes to sweep and display results. Selecting the data point larger than what
you need for a measurement will result in unnecessarily long sweep time. It is
recommended that you select high resolution data points only for an instance of
measuring wide frequency bands or requiring precise measurement data.
4.
Toggle the
Display
soft key and select
Magnitude
to view measurements in dB or
Phase
in
degree.
5.
Optional
. To turn the bias tee on, complete the following steps:
a.
Toggle the
Bias Tee
soft key and select
On
.
b.
Enter a value between 12 and 32 with 0.1 V step.
c.
Enter the
V
soft key.
Summary of Contents for CellAdvisor JD780A Series
Page 52: ...JD780A Series Analyzers 52 Getting Started Figure 6 Example of Web remote on Chrome...
Page 254: ...JD780A Series Analyzers 254 Using LTE FDD Signal Analyzer 3 RS SINR 4 S SS RSSI 5 P SS Power...
Page 323: ...JD780A Series Analyzers Using LTE TDD Signal Analyzer 323 3 RS SINR 4 S SS RSSI 5 P SS Power...
Page 390: ...JD780A Series Analyzers 390 Using EV DO Signal Analyzer 3 Pilot 4 MAC 64 5 MAC 128...
Page 525: ......
Page 590: ...JD780A Series Analyzers 590 Appendix Figure 306 Append keywords window...