JD780A Series Analyzers
462
Using Mobile WiMAX Signal Analyzer
Figure 230
Spectrum measurement screen with Mobile WiMAX signal analyzer
Setting trace
You can display up to six traces on the measurement chart simultaneously.
Procedure
1.
Press the
TRACE/DISPLAY
hard key.
2.
Press the
Select Trace
soft key, and then select the trace number:
T1
,
T2
,
T3
,
T4
,
T5
, or
T6
.
The legend shape of the selected trace changes from square to round to indicate that the trace is
the active one now.
3.
Do one of the following:
To
Select
Trace Legend
Clear current data and display with new
measurements
Clear Write
W
Display the input signal
’s maximum response only
(unlimited or for a certain amount of time)
Max Hold
M
Display the input signal
’s minimum response only
(unlimited or for a certain amount of time)
Min Hold
m
Capture the selected trace and compare traces
Capture
C
Load a saved trace
More (1/2) > Load
L
Hide the selected trace
Trace View > Off
F
Remove all the traces and initialize the trace
settings
More (1/2) > Trace
Clear All
NOTE
For the
Max Hold
and
Min Hold
, your instrument compares newly acquired data with the
active trace and displays larger maximum values or smaller minimum values on the
screen. You can set it to
Unlimited
to hold and view maximum or minimum data or
specify a certain amount of time up to 60 seconds by using numeric keys or the rotary
knob.
Summary of Contents for CellAdvisor JD780A Series
Page 52: ...JD780A Series Analyzers 52 Getting Started Figure 6 Example of Web remote on Chrome...
Page 254: ...JD780A Series Analyzers 254 Using LTE FDD Signal Analyzer 3 RS SINR 4 S SS RSSI 5 P SS Power...
Page 323: ...JD780A Series Analyzers Using LTE TDD Signal Analyzer 323 3 RS SINR 4 S SS RSSI 5 P SS Power...
Page 390: ...JD780A Series Analyzers 390 Using EV DO Signal Analyzer 3 Pilot 4 MAC 64 5 MAC 128...
Page 525: ......
Page 590: ...JD780A Series Analyzers 590 Appendix Figure 306 Append keywords window...