Chapter 4
Configuring the reflectometry test
Configuring the unit for SmartOTDR™
18
User Manual
770000102/08
Configuring the unit for SmartOTDR™
Once the SmartOTDR™ icon is validate, the
Results
page displays
automatically.
Before any test in SmartOTDR™:
1
select the configuration file, which contains all acquisition parame-
ters and file storage setup, and which has been created in Expert
mode (see
“Saving configuration parameters in a file” on page 33
).
2
the user can then configure some parameters before starting the
test
Selecting the
configuration
file
To load the configuration file to be used for SmartOTDR™ test:
1
Press
F
ILE
hard key.
2
In the
Explorer
page, on the left of the screen, select the storage
media and directory into which file configuration is saved.
3
Select the file on the right list: icon
/ type: .Config.
4
Press
Load
>
Load as SMART Config.
menu keys.
A beep is emitted to validate the selection of the configuration file.
5
Press
R
ESULTS
hard key to go to results page and perform OTDR
acquisition with the pre loaded configuration.
Fig. 9
Load file as SmartOTDR™ Configuration
NOTE
Some configuration files are available into the equipment, in
disk/
config
.