Chapter 10
Technical specifications
Technical specifications of the FiberComplete modules
150
User Manual
770000102/08
–
Signal including lambda information for the power meter (Auto
).
–
Modulated signal to identify the fiber (at 270 Hz/330 Hz/1kH/2kHz).
–
Wavelengths activated one after the other (TwinTest mode).
Technical specifications of the FiberComplete modules
Typical values, measured at 25°C unless specified.
For the OTDR specifications, please refer to
.
Bi-directional testing
Wavelength
1310 / 1490 / 1550 / 1625 nm
a
a. 1625nm not available with filtered version
Insertion Loss
Dynamic range
40 dB typ
IL uncertainty
+/- 0,25 dB
b
b. side by side reference
IL repeatability
< 0,05 dB
c
c. without disconnection
Display resolution
0.01 dB
ORL
ORL range
up to 55 dB
ORL uncertainty
+/- 0,5 dB
d
d. from 10 to 45 dB range
ORL repeatability
< 0,1 dB
e
e. from 20 to 40 dB
Display resolution
0.01 dB
Length
f
f. measurement @ 1550 nm with an index of refraction n = 1.468
Measurement range
150 km
Accuracy
- From 50 m to 20 km range
- Above 20 km
+/- 30m
+/- 100m
MTS2100 4100 Module US rev08.book Page 150 Jeudi, 1. août 2013 1:43 13