Maintenance and service

9.

NOTE

If an item is returned to JcT Analysentechnik, for main-

tenance or repair reasons, it will only be accepted after

the  RMA  form  on  our  website  has  been  completed

(www.jct.at/rma). This is to ensure the security of JcT

staff.

Recycling

The unit contains elements which are suitable for recy-

cling, and components which need special disposal. you

are therefore requested to make sure that the unit will

be recycled by the end of its service life.

Replacement of O-rings

9.1.

O-rings are consumables. Replace indurate or damaged

O-ring seals (A, B, c) of heated sample line or filter lock. 

Shut down JPES sample probe and heated sample

line and wait for cool down.

Pull off O-rings with non metallic tool (wood- or plas-

tic wedge). 

Apply a thin wetting of PTFE paste on O-rings and

pull them on.

Replacement of filter element

9.2.

Filter elements, O-rings and gaskets are consumables

and have to be replaced regularly, at least once a year.

Ensure that sealing surfaces are clean and unheart 
Please note that FFkM sealing materials are aging irre-

versibly  at  high  temperatures. This  process  is  called

“outgasing”.

Burn hazard!

Use heat resistant gloves.

CAUTION

The housing of the probe may get very hot!
Take care, in case of process over pressure, explosive

and/or toxic gas emanation is possible.
To avoid accidents take care for necessary safety pre-

cautions in case of service and maintenance.

Manual JPES

Wartung und Service

9.

HINWEIS

Ist es zu Wartungs- oder Reparaturzwecken notwendig,

das Gerät an JcT Analysentechnik zu schicken, ist das

RMA-Formular auf der Website vollständig auszufüllen

(www.jct.at/rma).  Andernfalls  kann  das  Gerät  zum

Schutz der JcT Mitarbeiter nicht übernommen werden.

Recycling

Das Gerät enthält Bauteile, die wiederverwertet werden

können, sowie Bauteile, die speziell entsorgt werden

müssen. Sorgen Sie deshalb dafür, dass das Gerät nach

der Verwendung der Wiederverwertung zugeführt wird.

Ersetzen der O-Ringe

9.1.

O-Ringe sind Verbrauchsteile. Bei Beschädigung oder

Verhärten der O-Ring Dichtungen (A, B, c) der beheiz-

ten  Messgasleitung  oder  des  Filterverschlusses  sind

diese zu ersetzten.

Dazu JPES Entnahmesonde und beheizte Mess-

gasleitung außer Betrieb setzten, spülen und abküh-

len lassen.
Die zu tauschenden O-Ringe mit einem nicht metal-

lischen Werkzeug (holz- oder kunststoffkeil) entfer-

nen.
Neuen Dichtring dünn mit PTFE Paste benetzen und

aufziehen.

Ersetzen des Filterelementes

9.2.

Filterelemente und Dichtungen sind Verbrauchsteile und

sind abhängig von den Einsatzbedingungen regelmäßig,

mind. 1mal pro Jahr zu warten. Es ist sicherzustellen,

dass die Dichtflächen sauber und unversehrt sind.
Es ist zu berücksichtigen, dass FFkM Dichtungsmate-

rialien bei hohen Temperaturen einem irreversiblen Al-

terungsprozess genannt “outgasing” unterliegen.

Verbrennungsgefahr!

hitzebeständige handschuhe benutzen.

VORSICHT

Das Gehäuse der Sonde kann sehr heiß sein!
Bei Prozessüberdruck können explosive und/oder toxi-

sche Gase austreten.
Entsprechende Maßnahmen sind bei Wartung und Ser-

vice sowie Ersetzen oder Reinigen des Filterelementes

zu treffen.

BA_DE_JPES_v1.5

––––––––––––––– 

[ 14 / 20 ]

–––––––––––––––––

Summary of Contents for JPES

Page 1: ...JCT ANALySENTEchNIk GMBh BEDIENUNGSANLEITUNG OPERATING MANUAL 02 18 JPES ...

Page 2: ...8 3 4 Pressure characteristics 8 3 5 Flow charts 9 4 Installation unpacking 9 5 Installation instructions 9 5 1 Mounting 9 5 2 Installation example 11 5 3 Electrical connections 11 5 4 Electrical interface 11 5 5 Electrical interface probe only 12 6 Start up 12 7 Demounting 13 8 Transport storage 13 9 Maintenance and service 14 9 1 Replacement of O rings 14 9 2 Replacement of filter element 14 10 ...

Page 3: ...e beheizte Gasentnahmesonde JPES ist für den mobilen Einsatz in Gasanalysesystemen bestimmt Beachten Sie die Angaben in den technischen Spezifi kationen hinsichtlich Umgebungs und Versorgungsbe dingungen sowie zulässige Temperaturgrenzen Introduction 1 The JPES series of heated gas sample probes is desig ned for use with dust and aerosol containing gases in extractive sampling systems especially f...

Page 4: ... Isolierung 4 Quick On Verriegelung 5 Beheizte Messgasleitung 6 kabeldose 7 Leiterplatte 8 Status und Netzanzeige 9 Tragegriff 10 O Ring B Filterverschluss aussen Ø 33 mm 11 O Ring A Filterverschluss innen Ø 15 mm 12 O Ring c Quick On Anschlüsse Ø 6 mm 1 Filter lock 2 Filter element 3 Isolation 4 Quick On lock 5 heated sample line 6 cable connector 7 Printed circuit board 8 Status and supply indic...

Page 5: ...to heated sample pipe 5 Quick on connector to sample probe 6 O ring c Quick On connectors Ø 6 mm 7 heater indicator 1 Quick On Sondenanschluss 2 Integrierter Temperatur Regler 3 0 3 m kabel mit kabelstecker von heizleitung JhSo 4 0 5 m kabel mit kabeldose zur Sonde 5 Doppelmantel Entnahmerohr 6 Gewindeanschluss R 3 8 7 O Ring c Quick On Anschlüsse Ø 6 mm 8 heizungsanzeige 9 Verschiebeflansch für Z...

Page 6: ...en 3 JPES 3 1 Technische Änderungen vorbehalten BA_DE_JPES_v1 5 6 20 Betriebsdaten Filterelemente GF PTFE Edelstahl Filteroberfläche 50 cm2 Arbeitsdruck 50 kPa Durchfluss bis zu 200 NL h filterabhän gig Gasberührende Materialien 1 4571 Viton Betriebstemperatur nominal 180 c Betriebstemperatur max 200 c Aufheizzeit 15 min Zulässige Umgebungstem peratur 20 c 55 c Schutzart IP 22 Einsatzort Nur für d...

Page 7: ...gnal Volt free contact contact load min 24V DAc 50 mA Low temperature contact limit max 230 VAc 5A cosP 0 95 Switching hysteresis 150 c 5 k connector 7 pin 15 k Beheizte Messgasleitung Anschlussspannung 115 VAc oder 230 VAc 50 60 hz Leistungsaufnahme ca 100 W m heizung Integrierter Temperatur Reg ler Außenschutz PA Geflecht Außendurchmesser 35 mm Minimaler Biegeradius 50 mm Farbe Schwarz Betriebst...

Page 8: ...ewicht 1 8 3 0 3 6 kg Betriebstemperatur 180 c Messgastemperatur max 250 c Aufheizzeit 15 min Anschlussstecker 7 polig Speisung über heizleitung Schutzart IP54 Heated sample pipe Supply voltage 115 VAc or 230 VAc 50 60hz Power consumption approx 530 W heater Integrated temperature controller Gas wetted parts SS 316 Ti Outer diameter 25 mm Nominal width inner pipe ID 4 mm Length 1000 1500 2000 mm L...

Page 9: ... 3 5 Installation Sichtkontrolle 4 Nach dem Auspacken ist das Gerät auf allfällige Trans portschäden zu untersuchen Wurde ein Schaden fest gestellt sind unverzüglich die verantwortliche Spedition und der händler zu benachrichtigen Es ist zu überprüfen ob die Gerätelieferung Ihrer Bestel lung entspricht Installationsvorschriften 5 Bei Arbeiten am elektrischen Teil des Gerätes ist es vom Netz zu tre...

Page 10: ...ase for mounting sample pipe Manual JPES Sonde mit Dichtung oder Universal Adapter unter Zuhilfenahme von PTFE Paste an Prozessflansch oder Entnahmeöffnung montieren siehe Installati onsbeispiel Der im Lieferumfang enthaltene Universal Adapter ist bis zu einer Rohrnennweite von 60mm passend Größere Durchmesser sind auf Anfrage erhältlich JPES Sonde mit Montagekette abhängen Einbauwinkel gemäß tech...

Page 11: ...2 Elektrischer Anschluss 5 3 Örtliche Netzspannung Netzfrequenz und Leis tungsaufnahme mit den Angaben auf den Typen schildern von Sonde beheizter Messgasleitung und eventuell beheiztem Entnahmerohr vergleichen Stromversorgungsanschluss herstellen und JPES Statuskontakt für die Betriebstemperatur am Analy sator anschließen Im Bedarfsfall ist in der Zufüh rung der Energieversorgung ein 2 poliger Ne...

Page 12: ...mage Manual JPES Elektrische Verbindungen nur Sonde 5 5 VORSICHT Dieses Gerät wird mit Netzspannung betrieben Beim Betrieb dieses Gerätes stehen zwangsläufig bestimmte Teile dieses Gerätes unter gefährlicher Spannung Im Betrieb kann das Gehäuse der Sonde sehr heiß wer den Durch Abnahme des Gehäuses werden heiße Teile zugänglich Bei jeglichen Arbeiten an der Sonde ist das Gerät abzuschalten die Abk...

Page 13: ... durchführen 4 Sicherstellen dass Gerät Filterraum und Filter sauber sind und sich keine Fremdkörper darin befinden 5 kontrolle aller Anschlüsse und Leitungseinfüh rungen auf festen Sitz VORSICHT Vor dem Einschalten ist sicherzustellen dass die am Gerät eingestellte Betriebsspannung und die Netzspan nung übereinstimmen 6 Energieversorgung der JPES Entnahmesonde einschalten Die gelbe LED leuchtet N...

Page 14: ...st es zu Wartungs oder Reparaturzwecken notwendig das Gerät an JcT Analysentechnik zu schicken ist das RMA Formular auf der Website vollständig auszufüllen www jct at rma Andernfalls kann das Gerät zum Schutz der JcT Mitarbeiter nicht übernommen werden Recycling Das Gerät enthält Bauteile die wiederverwertet werden können sowie Bauteile die speziell entsorgt werden müssen Sorgen Sie deshalb dafür ...

Page 15: ...recommended to purge the gas sampling probe diligently in heated condition Dimensions 10 All dimensions in mm Manual JPES Für den Ersatz der Filterelemente sind folgende Schritte vorzunehmen Elektrische Zuleitung abschalten und warten bis die 1 Sonde abgekühlt ist Durch Drehen des Filterverschlusses Pos 1 Filter 2 element entnehmen Filterelement Pos 2 vom Trägerelement Pos 1 3 abziehen Filtereleme...

Page 16: ...g 1x Entnahmerohr Ø 10 mm Länge 300 mm 3 1x Sample pipe Ø 10 mm length 300 mm 1x Filtereinsatz Mikrofaser 4 1x Filter microfiber 1x Beheizte Messgasleitung 3 m 5 1x heated sample line 3 m 1x Demontagehilfe 6 1x Disassembling aid 1x Transportkoffer 7 1x carrying case 1x Betriebsanleitung 8 1x Operating manual 1x Montagekette L 2 x 1 m 9 1x Mounting chain l 2 x 1 m 1x Entnahme Anschlussadapter Unive...

Page 17: ...Sample pipe Ø 8 mm length 300 mm 1x Entnahmerohr Ø 10 mm Länge 300 mm 6 1x Sample pipe Ø 10 mm length 300 mm 5x Filterelement Mikrofaser 7 5x Filter element microfiber 2x Filterelement PTFE 8 2x Filter element PTFE 1x Beheizte Messgasleitung 5 m 9 1x heated sample line 5 m 1x O Ring Satz 10 1x O ring set 1x PTFE Paste 113 4 g Tube 11 1x PTFE paste 113 4 g collapsible tube 1x Demontagehilfe 12 1x D...

Page 18: ...erohre mit integriertem Temperaturregler Heated pipes with integrated temperature controller 34 00605 beheiztes Entnahmerohr 230 VAC L 0 5 m heated sample pipe 230 VAC l 0 5 m 34 00610 beheiztes Entnahmerohr 230 VAC L 1 m heated sample pipe 230 VAC l 1 m 34 00615 beheiztes Entnahmerohr 230 VAC L 1 5 m heated sample pipe 230 VAC l 1 5 m 34 00620 beheiztes Entnahmerohr 230 VAC L 2 m heated sample pi...

Page 19: ... case yellow with shell moulds Art Nr Part No Beschreibung Description 34 90011 O Ring A Filterverschluss innen Ø 15mm O ring A Filter lock inside Ø 15mm 34 90013 O Ring B Filterverschluss außen Ø 33mm O ring B Filter lock outside Ø 33mm 34 90012 O Ring C Quick On Anschlüsse Ø 6mm O ring C Quick On connectors Ø 6mm 34 90010 O Ring Satz bestehend aus 1x O Ring A 1x O Ring B 4x O Ring C O ring set c...

Page 20: ...ax 43 2622 87 2011 sales jct at Firmenbuchgericht kreisgericht Wr Neustadt FN 119682 d UID Nummer ATU 20559900 Manual JPES BA_DE_JPES_v1 5 20 20 Mehr Informationen zur JPES finden Sie auf unserer Website Please refer to our website for more information on the JPES ...

Reviews: