General safety rules
General
The system is for INDOOR use only and is designed to be mounted
on a medical cart.
Switch off the central power switch and unplug the mains socket
before moving. Make sure the mains cable is secured on the cart
while moving the system.
The unit is equipped with ventilation slots. NEVER COVER THESE
SLOTS AND AVOID LIQUIDS OR (METAL) OBJECTS FROM GETTING IN.
If fuses fail, switch off the system and remove the power plug
before replacing them. Replace only by fuses of the same type
and current.
Make sure the batteries are in optimum condition at all times by
means of charging them when the system is not in use.
Do not connect more than 4 medical devices to the system in
order to keep the earth leakage current on the output below
5mA.
The supplied 5 meter long power cord is rated for medical use.
Always connect power plug to a medical grade outlet to ensure
grounding protection, and always connect the equipotentiality
plug to a valid equipotentiality terminal before use.
Inspect the power cord before each use. DO NOT USE CORD IF
DAMAGED.
DO NOT plug more than the specified number of watts into socket
strip.