Step 5: Adjust the Image Quality
Related Setting Items:
To maximize the performance of the camera, configure its basic function in the following order.
Note:
On this camera, you must configure both the RGB and SWIR channels individually.
Configure the Line Rate
You can set the line rate to 1L or more.
This function can be used to match the scanning speed of the camera to the feeding speed of the
object or to lengthen the accumulation time to increase sensitivity.
Note:
Refer to the
and
Control Without External Triggers
topics.
DSNU Correction (Pixel Black Correct)
DSNU (dark signal non-uniformity) is a variation between pixels in the dark areas generated by
the sensor. If the line rate is slowed or a long exposure time is set, the dark current in the sensor
may change and the state of the DSNU may change.
RGB
1. Specify the user area (User1 ~ User3) to save the black level correction value with
PixelBlackCorrectionMode
.
Note:
You cannot perform calibration when
Off
or
Default
is selected.
2. Black level correction data is automatically generated by
PerformPixelBlackCalibration
and saved in the user area specified in step 1.
3. You can check the execution result of black level correction by
PixelBlackDetectResult
.