Chapter 5
– Setup Menu
53
M
ODE
F
IELD
Two options are available in this field:
Match Sample and Match Data.
In the
Match Sample
mode, white point calibrations are based on a measured reference display
– “Golden
Sample”.
Match Data
mode allows the user to enter target values in terms of luminance (Y) and CIE 1931 x, y values.
Whichever method is chosen, all
R
G
B
measurements require that a
Phosphor
(either stored or custom) be
selected and that a reference white point is set.
S
ELECTING A
P
HOSPHOR
Proper white point calibration is dependent on the color characteristics of the display primaries. Phosphor is
used as a historical reference to CRT’s. The data for several types of phosphors are pre-stored in the PR-7XX.
They are CIE, NTSC, EBU (PAL / SECAM), SMPTE and HDTV. The user may also enter display primary
chromaticity values or use measured values.
1. To select a pre-stored phosphor set touch
Select
located to the right of the
Phosphors
window. The
following screen appears:
FIGURE 46 -
R
G
B
PHOSPHOR SELECT SCREEN
Summary of Contents for Photo Research SpectraScan PR-7 Series
Page 2: ......
Page 10: ...PR 7XX User s Manual vi...
Page 32: ......
Page 84: ......
Page 91: ...Chapter 7 File Settings 87 FIGURE 73 ILLUMINANCE LUMINANCE CONCEPT...
Page 92: ......
Page 102: ......
Page 108: ......
Page 117: ...Chapter 12 Connectivity 113...
Page 118: ......