Chapter 5
– Setup Menu
45
2.
Enter the number of measurements to average. The acceptable range is 1 to 99.
3.
Touch the
Back
key when done.
H
OW TO
S
ET THE
A
DAPTIVE
E
XPOSURE
T
IME
The exposure time is the amount of time the detector is exposed to light. The PR-7XX utilizes the patented
Adaptive Sensitivity
algorithm that automatically selects the proper exposure time for the available signal
insuring the most accurate measurement for the available light. Use the following procedure to set Adaptive
Sensitivity to ON.
1.
To set the instrument to
Adaptive Sensitivity
, make sure the field next to
Adaptive
reads
Enabled.
H
OW TO
S
ET A
F
IXED
E
XPOSURE
T
IME
It may be desirable to set a fixed exposure time for an application. For example, when measuring a display for
luminance uniformity, the exposure time will be the virtually identical for each measurement since the luminance
levels are very similar for every part of the display. In this example, using Adaptive Sensitivity may make the
measurement time longer because of the overhead the algorithm requires in its iterative process to find the ideal
exposure time. Use the following procedure to set a fixed exposure time.
1.
If the field immediately to the right of
Adaptive
reads
Enabled
, touch the field to set
Adaptive
to
Disabled
.
2.
Touch the field adjacent to
Exposure
to display the following data entry menu:
FIGURE 38
– EXPOSURE TIME ENTRY SCREEN
3.
Enter the desired exposure time in milliseconds (ms). The range is 7 ms to 120,000 ms (2 minutes) for
Standard Sensitivity and 7 to 300,000 ms (5 minutes) for Extended Sensitivity.
Note: The sensitivity range must be set to Extended to use the maximum value of
300,000 ms (5 minutes).
4.
Touch
Back
when done.
H
OW TO
S
ET THE
I
NSTRUMENT
S
ENSITIVITY
R
ANGE
Summary of Contents for Photo Research SpectraScan PR-7 Series
Page 2: ......
Page 10: ...PR 7XX User s Manual vi...
Page 32: ......
Page 84: ......
Page 91: ...Chapter 7 File Settings 87 FIGURE 73 ILLUMINANCE LUMINANCE CONCEPT...
Page 92: ......
Page 102: ......
Page 108: ......
Page 117: ...Chapter 12 Connectivity 113...
Page 118: ......