Appendix 1
– Remote Control Mode
134
R
EMOTE
C
ONTROL
E
RROR
C
ODES
R
EMOTE
C
ONTROL
M
EASUREMENT
E
RRORS
Error
Meaning
-0001
Light source not constant.
-0002
Light overload
– signal too intense.
-0003
Cannot Sync to light source. Light source frequency below 20Hz, above 400
Hz or signal too low to Sync.
-0004
Adaptive mode error.
-0008
Weak light
– insufficient signal.
-0009
Sync Error.
-0010
Cannot Auto Sync to light source.
-0012
Adaptive mode time out. Light source not constant.
R
EMOTE
C
ONTROL
P
ARSING
E
RRORS
Error
Meaning
Valid Values
-1000
Illegal command
-1001
Too many fields in setup command
-1002
Invalid primary accessory code
-1003
Invalid Addon 1 accessory code
-1004
Invalid Addon 2 accessory code
-1005
Accessory is not a primary accessory
-1006
Accessory is not an Addon accessory
-1007
Accessory already selected
-1008
Invalid Aperture index (PR-7XX only)
-1009
Invalid units code
0 = English
1 = Metric (SI)
-1010
Invalid Exposure value
3 to 6000 ms
PR-7XX
6 to 30,000 ms
-1011
Invalid Gain code
0 = Normal
1 = 1X for AC sources
2 = 10X
3 = 100X
-1012
Invalid average cycles
1 to 99
-1013
Invalid Calc Mode
-1014
Invalid Trigger Mode
-1015
Invalid CIE observer
2 or 10
Summary of Contents for Photo Research SpectraScan PR-7 Series
Page 2: ......
Page 10: ...PR 7XX User s Manual vi...
Page 32: ......
Page 84: ......
Page 91: ...Chapter 7 File Settings 87 FIGURE 73 ILLUMINANCE LUMINANCE CONCEPT...
Page 92: ......
Page 102: ......
Page 108: ......
Page 117: ...Chapter 12 Connectivity 113...
Page 118: ......