PhaseStation 53100A User’s Manual V1.04 25-Apr-2022
189
Note that the accuracy of the absolute frequency readings obtained with the
/format:F
switch depends only on the reference source, regardless of the nominal input frequency
reported by
show inputs
. Consequently, it is unaffected by the
/inputfreq
or
/roundfreq
options. As a result, the
/inputfreq
and
/roundfreq
options are not required in most
TSERVE measurements. However, note that if the nominal reference frequency is not a
multiple of 100 kHz, it is necessary
to use
/referencefreq
to avoid inaccurate results
regardless of the selected output format. There is no corresponding way to increase the
reference frequency’s rounding precision; the nominal reference frequency must be
specified explicitly with
/referencefreq
if 100-kHz precision is insufficient.
/ sep:<character>
/ timestamp:<s, MJD, UNIX>
These related options cause TSERVE to add a timestamp to each phase-difference or
frequency reading that is written to a shared file or streamed to network clients.
By default, no timestamps are emitted. When
/timestamp:MJD
is specified, the
Modified Julian Date (MJD) associated with each phase-difference or frequency value
will be written as the first numeric field on each line, followed by a space character
(ASCII 32), then the measurement value itself. MJD timestamps are generated with six
digits of decimal precision, resulting in roughly 10 unique MJD values per second.
You can also use
/timestamp:s
to emit a simple numeric timestamp at one-microsecond
precision. These timestamps begin at 0.000000 for the first reading acquired. Finally,
/timestamp:UNIX
can be used to request numeric timestamps at one-millisecond
precision relative to the standard Unix epoch (January 1
st
, 1970 at 00:00:00 UTC.)
In all cases, the difference between successive timestamps will be equal to the
reciprocal of the
/phaserate
parameter, whose default is 100 readings per second,
scaled further by any specified
/phasedec
ratio beyond its default value of 2:1. Refer to
the description of the
/phaserate
parameter for more information.
Summary of Contents for PHASE STATION 53100A
Page 1: ...TM 53100A Phase Noise Test Set User s Manual Revision 1 04 April 25 2022...
Page 16: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 16...
Page 18: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 18...
Page 20: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 20...
Page 30: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 30...
Page 36: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 36...
Page 42: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 42...
Page 64: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 64...
Page 74: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 74...
Page 104: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 104...
Page 116: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 116...
Page 122: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 122...
Page 130: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 130...
Page 140: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 140...
Page 214: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 214...
Page 215: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 215 Appendix Legal and regulatory notices...