LC_XO GPSDO User Manual
36
© 2012 Jackson Labs Technologies, Inc.
A health status of 0x0 indicates a properly locked, and warmed-up unit that is completely healthy.
4.5.10 SYNChronization?
This query returns the results of these six queries :
SYNChronization:SOURce:MODE?
SYNChronization:SOURce:STATE?
SYNChronization:LOCKed?
SYNChronization:HOLDover:DURation?
SYNChronization:FEEstimate?
SYNChronization:TINTerval?
SYNChronization:health?
4.6 DIAGnostic Subsystem
This subsystem regroups the queries related to the diagnostic of the OCXO.The list of the commands
supported for this subsystem is as follows:
DIAGnostic:ROSCillator:EFControl:RELative?
DIAGnostic:ROSCillator:EFControl:ABSolute?
4.6.1 DIAGnostic:ROSCillator:EFControl:RELative?
This query returns the Electronic Frequency Control (EFC) output value of the internal reference
oscillator. It returns a percentage value between -100% to +100%. :
4.6.2 DIAGnostic:ROSCillator:EFControl:ABSolute?
This query returns the Electronic Frequency Control (EFC) output value of the internal reference
oscillator. It returns a value in volts between 0 and 5 V
4.7 MEASURE Subsystem
The MEASURE commands are not supported on the LC_XO board.
4.7.1 MEASure:VOLTage?
This command is not supported on LC_XO, and will return undetermined values.
4.7.2 MEASure:CURRent?
This command is not supported on LC_XO, and will return undetermined values.
Summary of Contents for LC_XO
Page 2: ...Revision 08 29 12 B LC_XO PINOUT DESCRIPTION...
Page 3: ...LC_XO GPSDO User Manual Document 80200517 Version 1 0 Date 23 September 2012...
Page 4: ...LC_XO GPSDO User Manual Copyright 2012 Jackson Labs Technologies Inc...
Page 8: ...LC_XO GPSDO User Manual iv 2012 Jackson Labs Technologies Inc...
Page 36: ...LC_XO GPSDO User Manual 28 2012 Jackson Labs Technologies Inc...