82
©2023 J.A. Woollam Co.
M-2000® Hardware Manual
Figure 4-55. Automatic Sample Alignment options.
9.
The Camera Detector is now calibrated to the system alignment. When
samples are mounted, this Lookdown Detector can be trusted to give the
correct sample alignment. After this point, all sample measurements
may use automated alignment.
Note:
If a manual tilt sample stage (example: Rotator), the Tilt Camera
may still be used for sample tilt positioning – cross hair position
4.7
Sample Alignment (with
Manual Tip/Tilt)
Note:
Automated Tip/Tilt alignment is an optional accessory that may or
may not have been purchased with your system. If your purchase
included Automated Tip/Tilt, please see Section 4.8 for Sample
Alignment (with Automated Tip/Tilt).
Note:
This procedure must be performed every time a sample is mounted
on the ellipsometer, prior to measurement or calibration.
2.
Place the sample on the sample stage.
3.
Select
Hardware
>
Controls
>
System
>‘Angle’. Type 65 in the dialog box
and press ‘Ok’.
Figure 4-56. Move Angle dialog box.