The J.A. Woollam Co. M-2000 ellipsometers are low maintenance instruments.
The only required maintenance is to change the lamp(s) as needed and to check
the alignment of the system.
10.1
D2 & QTH Lamps (M-2000D,
DI, U, & UI) in XLS-200
The lamps must be replaced after a certain period of time. As a lamp
approaches the end of its useful life, intensity levels can fluctuate and spectral
intensity can be reduced, both may result in poor or noisy data.
LAMP
VENDOR
MODEL#
COLOR
TEMP.
RATING
TYPICAL
LIFETIME
QTH
ILT
L9404
2900°K
2000
hours
3000 hours
D2
Hamamatsu L10804
N/A
2000
hours
3000 hours
Table 10-1. Lamps and typical lifetimes.
Lamp Installation/Replacement
1.
Turn off the M-2000 Detector and Lamp Power Supply Module. Allow
lamp(s) to cool for 2 hours.
2.
Remove the (2) M3 SHC screws (2.5mm Allen wrench) and remove the
cover on the light source.
3.
Put on latex gloves and safety goggles.
Warning: The lamps get extremely hot during operation. Always allow them
to cool adequately before replacing.