1.
The first prompt is to attach the Source probe.
Figure 6-12. Adjust Source probe.
2.
Using the three captive thumb screws, attach the Source focusing probe
to the Source probe tilt stage. Click OK when finished.
Figure 6-13. Focus probe mounting.
3.
Carefully center the cross hair by translating the Source side lens with the
Vertical Source (VS) and Horizontal Source (HS) adjustment screws. Close
alignment screen when finished.
Note:
If the beam is not entering the Receiver Unit at all, ‘Out of Range’
will be displayed.