3.3.8 Setup
The following steps will guide you through a basic setup for turning your system on for the first time:
CAUTION
The Motor/Base Module number selection in the Setup C screen MUST match the Motor/Base Module
attached to the Controller Module. An incorrect setting could overheat the motor and damage the equipment
and cause a hazardous condition.
1. On the rear of the Controller Module
a. Make sure the 1/0 (On/Off) switch is set to 0 (Off).
b. Connect the power cord to the Controller Module and the power source.
c. Connect the Motor Cable.
2. On the Motor/Base Module
a. Connect the other end of the Motor Cable
b. Set the AP Style displacement adjust to 4.5 or
c. Set the Heavy Duty Style displacement adjust to 10
3. On the Pump Module
NOTE
Make sure the inlet tubing is larger than, or the same size as, the outlet tubing. Start with the reservoir even
with or slightly higher than the Pump Module and the Pump Module even with or slightly higher than the
dispense tip. Adjust as necessary to fit your application.
a. Connect the inlet fitting and tubing
b. Connect the gland inlet fitting and tubing (
*
optional)
c. Connect the outlet fitting and tubing
d. Connect the gland outlet fitting and tubing (
*
optional)
3.3.9 Start-up
Switch the 1/0 power switch to the '1' position. The Power Up screen will appear and the Idle indicator will illuminate.
This screen displays the firmware version and allows access to the Prime screen and Dispense or Meter screen.
Refer to Section 3.2.1 for a description of each field. Go to the Dispense section (3.2.9.2) or Meter section (3.2.9.3 if
the system has previously been primed. If not, follow the instructions in the following section.
NOTE
All controllers are sent in Supervisor Level.
3.3.9.1 Prime
Press the PRIME SCREEN push-button to enter Prime mode.
Position the pump for priming (outlet above or even with the inlet) fill all reservoirs with liquid, and make sure all liquid
lines are connected and secure. Press the START push-button to begin priming. The system will prime based on the
Priming Rate and Prime Volume settings. Pressing the STOP push-button can stop the priming any time.