Page 1: ...Quick Infusion Set Up Guide Ivac 572 Huw Evans B Sc HND Eng EBME ...
Page 2: ...________________________________________________ 1000ml to be infused in 8 hour 30 minutes 100ml hr Again change the time of infusion into minutes 8 hour 30 minutes 510 minutes to calculate this rate 60 x 1000 117 6 ml hr 510 The rate of infusion is therefore 118 ml hr Before you start check that your rate calculations are correct Try these and see how you get on Quick Start Up Guide Setting up an...
Page 3: ...ntil the power up sequence has finished The TOT VOL INF can be re set at this point by pressing the TOT VOL INF key along with the 100 s down arrow as shown on the pump to clear circled in YELLOW This will clear all the previous total fluid volumes Setting up an infusion for the Ivac 572 Huw Evans B Sc HND Eng EBME ...
Page 4: ...osite ensuring that the pressure sensor is securely fitted and that the line passes over the AIL detector circled in RED During an AIL error code clean the sensor this may clear the error code Setting up an infusion for the Ivac 572 Huw Evans B Sc HND Eng EBME ...
Page 5: ...ed on the clear part of the drip chamber Ensuring that the flow sensor is clean will lessen some error codes Press Rate Key key 11 and set rate to 155ml by pressing the UP DOWN arrows shown opposite circled in RED Setting up an infusion for the Ivac 572 Huw Evans B Sc HND Eng EBME ...
Page 6: ...structions until infusion begins Once the infusion begins ensure that the flow indicator bar circled in blue opposite rotates during infusion and that it will change from a rotating bar to a circle Each circle represents a drip being detected Setting up an infusion for the Ivac 572 Huw Evans B Sc HND Eng EBME ...
Page 7: ...READ PRES key circled in YELLOW Once the infusion has completed it will alarm and the display screen will show INFUSION COMPLETE KVO 5 0ml HR The KVO Keep Vein Open indicates that the pump will continue to infuse at 5ml hl to stop the KVO press the hold key Setting up an infusion for the Ivac 572 Huw Evans B Sc HND Eng EBME ...
Page 8: ...off by pressing the On Off key Before storing the pump discard the giving set safely as per hospital policy Ensure that the pump is clean before storing Always ensure that the pump is plugged into the mains socked so that the pump is continually charging Stopping an infusion Huw Evans B Sc HND Eng EBME ...