
Compound Control
Command
Copyright © Itech Electronic Co., Ltd.
26
Chapter8 Compound Control Command
APPLy {<
voltage value
>|DEF|MIN|MAX} [{<
current
value
>|DEF|MIN|MAX}]
This order has combined two kinds of orders: VOLTage and CURRent. When
sending this order to unit, power supply will output voltage and current
according to the current setting of this order. Of course the precondition is the
setting value is within the set range. If not, then a execution error will occur. You
can also use DEF
、
MIN or MAX as the setting parameters.DEF represents
factory setting. MIN will enable the voltage and current to be 0. MAX will enable
the output to be the highest value in the range.
Order syntax
[SOURce:]APPLy <NRf>
Parameters
MIN to MAX
Unit
V, A
Query syntax
APPLy?
Return parameters
<NR2>