5.0 Troubleshooting
FAULT MESSAGE
CAUSE
CORRECTIVE
RELATED
OR CONDITION
ACTION
PROCEDURE
SECTIONS
CHARGE
Battery is discharged
Charge battery for 16 hours.
6.1, 7.0, 8.1, 8.2,
BATTERY
or fuse blown.
If message persists, check
9.0, 10.0
fuses and/or replace battery.
SAMPLE TOO
Excess or insufficient
Repeat test with new cuvette.
6.2, 7.0, 8.3, 8.4,
LARGE (STL)/
sample. Note: 15 ± 1.5 µL
If problem persists, a pump
8.5, 9.0, 10.0
SAMPLE TOO
of blood is aspirated by
may need to be replaced.
SMALL (STS)
the instrument.
PREMATURE
Sample was detected at front
Repeat test with new cuvette.
6.2, 7.0, 8.3, 8.4,
SAMPLE
detector before specified
If message persists, a pump
8.5, 9.0, 10.0
time period. May occur if
may need to be changed.
sample is added before
Contact ITC Technical Service.
pump-priming sequence
is complete.
DOES NOT ACCEPT
Key switch is not functional.
Remove and replace
6.3, 7.1, 9.0, 10.0
KEYPAD ENTRY
permanent or intermittent
membrane switch panel
and related assembly.
CODE
Photo System Error
Remove and replace LED
8.6, 10.0
Harness Assembly
HEMOCHRON
®
Jr. Signature Service Manual
6
Summary of Contents for HEMOCHRON Jr. Signature
Page 1: ...SERVICE MANUAL...
Page 15: ...11 0 FIGURES HEMOCHRON Jr Signature Service Manual 15...
Page 16: ...HEMOCHRON Jr Signature Service Manual 16 FIGURE 1...
Page 19: ...HEMOCHRON Jr Signature Service Manual 19 FIGURE 4 CONNECTOR J6...
Page 21: ...HEMOCHRON Jr Signature Service Manual 21 FIGURE 6 LED HARNESS PCB MOUNTING SCREWS...
Page 23: ...HEMOCHRON Jr Signature Service Manual 23...
Page 24: ...HJ7430 1 03 HEMOCHRON is a registered trademark of ITC...