APPENDIX 1: TROUBLE SHOOTING
TEMPERATURE UNSTABLE
Controller has incorrect parameters set; reset to standard. If still unstable try the self-tune option. Check the
electricity supply is stable.
UNIT FAILS TO HEAT
Check controller is set correctly. If the output indicator on the controller is lit the relay should operate the heater. If
the output indicator stays off it is likely the controller is set incorrectly.
CONTROLLER DISPLAY UNUSUAL
Error or system message. Refer to Isotech or your local Agent.
HEATER FAIL ON
If the temperature of the block increases rapidly with the controller output staying off the relay and sensor are
suspect. Do not use the unit, contact Isotech or your agent.
WARM UP TIME UNUSUALLY LONG
Controller set incorrectly, failure of one heater, or cooling fan permanently on. Check controller settings.
The fan on the side panel should always be on.
INDICATOR MISBEHAVES ('S' ONLY)
Wrong configuration code entered, check against table.
If SnSr FAIL is displayed, check probe connections. If displayed with no probe connected, check that the link is fitted
correctly and that the configuration is for a thermocouple type.
MAINTENANCE
The only moving part is the fan. It has sealed-for-life bearings. Depending on the environment in which it is used,
periodic cleaning of them and the inside of the case is recommended. Cleaning may be accomplished by the use of a
small dry paint brush.
The instrument should be periodically checked to ensure it is in good order both mechanically and electrically.