METHOD
Evaluation of the Optical Bench
The Pegasus 'R' item was set up in the 'normal' position as described above. The controller temperature was set to
1000 C. After stabilisation the 3 freedom of movement were evaluated.
a.
Back & Forth
The optical pyrometer was moved back & forth by 100mm. The following readings were obtained.
OPTICAL PYROMETER READING
650mm from rear of target to optical pyrometer
750mm
850mm
984
983
983
b.
At a distance of 750mm. The focal length of the optical pyrometer was changed.
METRES FOCAL LENGTH
READING
.75
.8
.85
.9
983
983
983
983
c.
The height of the optical pyrometer was altered
READING C
-2mm
Normal
+2mm
982
983
982
d.
The lateral position of the optical pyrometer was adjusted and the optical pyrometer rotated to align with the
centre of the target.
LATERAL MOVEMENT
READING C
Right (mm)
5
10
Left (mm)
5
10
983
983
983
983