SPECIFICATIONS
NHR Standard
NHR High Precision
Case
1/12 NIM standard cassette
Operating temperature
0 – 40 °C
Storage temperature
-20 − 60 °C
Humidity
20 – 80 %, not condensing
Notes:
1)
The resolution of measurable values depends on the settings of the sampling rate and the digital filter!
2)
Standard factory settings
Table 1: Technical data: Specifications
CONFIGURATIONS NHR
Type
Ch Precision V
nom
I
nom
Ripple (mV
p-p
)
HV output
mode
(V
mode
/ I
mode
)
Item Code
Options
>1kHz
10Hz-
1kHz
0.1Hz-
10Hz
NHR 20 20 2
Standard 2000 V
6 mA 3
10
n/a
2 kV / 6 mA
NR020020r605
VCT, IU, ID
NHR 20 60 2
Standard 6000 V
4 mA 3
10
n/a
6 kV / 2mA
4kV / 3mA
2kV / 4mA
NR020060r405
VCT, IU, ID
NHR 40 20 4
Standard 2000 V 6 mA 3
10
n/a
2 kV / 6 mA
NR040020r605
VCT, IU, ID
NHR 40 60 4
Standard 6000 V
4 mA 3
10
n/a
6 kV / 2mA
4kV / 3mA
2kV / 4mA
NR040060r405
VCT, IU, ID
NHR 22 20 2
High
2000 V 6 mA 2
2
5
2 kV / 6 mA
NR022020r605
VCT, IU, ID, TC, L
NHR 22 60 2
High
6000 V
4 mA 2
3
10
6 kV / 2mA
4kV / 3mA
2kV / 4mA
NR022060r405
VCT, IU, ID, TC, L
NHR 42 20 4
High
2000 V 6 mA 2
2
5
2 kV / 6 mA
NR042020r605
VCT, IU, ID, TC, L
NHR 42 60 4
High
6000 V
4 mA 2
3
10
6 kV / 2mA
4kV / 3mA
2kV / 4mA
NR042060r405
VCT, IU, ID, TC, L
Table 2: Technical data: Configurations
NHR – NIM standard High Voltage Power Supply | Last changed on: 25.11.2019 |
www.iseg-hv.com
10/31