iRhythm Zio XT Start Manual Download Page 1

START GUIDE

 

PLAN & POSITION

1

3

1

2a

2b

4c

4b

4a

Enroll the patient.

•  Online: 

www.zioreports.com

Write in Patient ID Number and Initials, Physician, Start Date & Time 
on cover of 

Patient Instructions & Button Press Log.

Determine position:

•  Patient should stand with arms relaxed by their sides.
•  Skin should be free of perspiration.
•  Determine placement area without removing backing.
•  Place on flattest part of upper left chest:

 

— 1 finger width below collarbone, centered over left  

pectoral muscle

 

— Edge of the Zio XT Patch next to sternum

 

— Angle so arrow on top label points upward

 

— Avoid armpit and breast tissue

Prepare a large area:

•  SHAVE

 area.

•  ABRADE

 skin, applying pressure for 40 broad strokes— 

10 times in the direction of each arrow in 2a & 2b 
(important for adhesion and ECG signal quality).

•  CLEAN

 skin thoroughly with all alcohol pads. Let dry for 1 minute.

•  Hold device in the center and remove clear backings. 

Keep top label on. Do not touch adhesive.

•  Place Zio XT Patch in the recommended position.
•  Press firmly across the entire device for 2 minutes (see image 3).

•  Peel off the 2 parts of the top label (see image 4a).
•  Press firmly across the entire device for 2 minutes, working  

adhesive into the skin. Emphasize pressing on the dotted areas  
shown in figure 4b.

•  Firmly press the ‘Zio’ button and release (see image 4c). The 

green

  

light will flash 5 times indicating that the monitoring has started.

•  Help patient get familiar with the location of the button.

 

PREP SKIN

 

2

 

APPLY PATCH

 

3

 

FINISH & ACTIVATE

 

4

Reviews: