iRhythm ZIO XT PATCH Quick Manual Download Page 1

3

3

4

2

1

 

PLAN & POSITION

Enroll the patient.

•  Online:  

www.zioreports.com

•  Phone:  

1-888-693-2401 

Write in Name, Physician, Start Date & Time on cover of 

Patient 

Instructions & Button Press Log

.

Determine position:

•  Patient should stand with arms relaxed at side.
•  Skin should be free of perspiration.
•  Determine placement area without removing backing.
•  Place on flattest part of upper left chest:

— 1 finger width below collarbone, centered over left 

pectoral muscle

— Edge of the ZIO® XT Patch next to sternum
— Angle so arrow on top label points upward
— Avoid armpit and breast tissue

 

PrEP SKIN

Prepare a large area:

 

SHAVE

 area if hair is present.

•  ABRADE

 skin, applying pressure for 40 broad strokes— 

10 times in the direction of each arrow in 2a & 2b 

(important for adhesion and ECG signal quality).

•  CLEAN

 skin thoroughly with both alcohol pads. Let dry for 

1 minute.

 

APPLY PATCH

•  Hold device in the center and remove clear backings. 

Keep top label on. Do not touch adhesive.

•  Place ZIO® XT Patch in the recommended position.
•  Press firmly across the entire device for 2 minutes (see 

image 3).

 

FINISH & ACTIVATE

•  Peel off the 2 parts of the top label (see image 4a).
•  Press firmly across the entire device for 2 minutes, working 

adhesive into the skin. Emphasize pressing on the dotted 

areas shown in figure 4b.

•  Firmly press the ‘ZIO’ button and release (see image 

4c). The 

green

 light will flash 5 times indicating that the 

monitoring has started.

•  Help patient get familiar with the location of the button.

Start Guide

1

2a

4c

2b

4b

4a

Reviews: