RS-485 Digital Communications
Modline 5
117
6.10
Command Code Summary
Code Command
Description
Type Group
Notes
AA
Analog alarm output
R/W
Analog
Out
Sets the Analog Output to a specified value when unit is in any of
two alarm conditions. (Error Codes X107, X109).
AC
Auto cal time period
(in hours)
R/W
Eng.
Sets the number of hours between Automatic Checks or performs
immediate check.
AF
Analog full scale
R/W
Analog
Out
Analog Full Scale command specifies the measured temperature
that will produce the Analog Output full scale current.
AH
Analog output high
temperature alarm
R/W
Analog
Out
Sets alarm condition at the Analog Output if the measured
temperature is above the value for Analog Full Scale Output.
AL
Analog output low
temperature alarm
R/W
Analog
Out
Sets alarm condition at the Analog Output if the measured
temperature is below the value for Analog Zero Scale Output.
AO
Analog output mode
R/W
Analog
Out
Sets the Analog Output Current loop range to either 0 to 4 mA or 4-
20 mA.
AT
Attenuation
Alarm threshold
R/W Eng.
Set the percentage value for the attenuation alarm threshold.
AZ
analog zero scale
R/W
Analog
Out
Sets the measured temperature that produces the Analog Output
Zero Scale current.
BT
Report Isoblock
temperature
R/O
Eng.
Returns the infrared detector isoblock environment temperature in
°C.
CL
Set to 1 or 2 color
mode
R/W
Head
Selects One Color or Two Color operation for a Series 5r ratio
Sensor.
DR
Peak picker decay rate R/W
Head
Sets the peak picker decay rate in F or C. Zero (0) sets the unit to
never decay.
DT
dirty window threshold R/W
Eng.
Dirty Window Detector Sensitivity Warning Setting
EM
Emissivity
R/W
Head
Sets Emissivity for single color or ratio units in one color mode
ES
E slope
R/W
Head
Sets the E-Slope for Ratio Sensors in the 2 color ratio mode
FT
Features Matrix
R/O
Eng.
Returns bit values for the features the Sensor supports.
KL
Keyboard lock
R/W
Eng
Locks or unlocks the Sensor’s rear panel keyboard.
LS
Laser Control
R/W
Eng
Turns the sighting laser on or off.
MD
Model
R/O
Eng
Returns the Model number of the Sensor.
MT
Match Temp Display
R/W
Head
Writes temperature value to be displayed to match target.
Read replies Emissivity or E-Slope value
PD
Peak delay
R/W
Head
Sets the delay time before Peak Picker action starts.
PK
peak picker reset
below
R/W
Head
Sets the temperature at which Peak Picker action starts with Auto
Peak Picker Reset Function
PR
Peak and reset
write Head
Write Only Command, Resets the Peak Picker
PS
Peak picker auto reset R/W
Head
Selects Peak Picker Auto Reset operation
RC
Case temperature
R/O
Eng
Returns Sensor’s internal Case temperature.
RP
Relay polarity
R/W
Eng
Sets the polarity of the alarm relay.
RR
Alarm Operation
R/W
Eng
When this is turned on, it also enables the attenuation alarm.
RT
Response time
R/W
Head
Sets the Sensor Response Time.
SG
Signal Conditioning
R/W
Head
Turns Peak Picker or Track and Hold on or off.
SN
Serial Number
R/O
Eng.
Returns the serial number string of the unit.
ST
System alarm status
R/O
Alarm
Sensor returns bit mapped values for alarms and the current
measurement condition status.