1-5
Figure 1-2.
1.4.6 Flow Test (3860+ or 3861 SideCar)
Connect patient end of IV Set to Infutest 2000 (or other IV Pump Flow
test device). Set Pump Rate to 100 ml and VTBI to 25 ml, verify Ready
to Start is flashing in lower right of the display. Press the Start/Stop
Channel A (Channel B for SideCar) key and verify the pump begins
running as indicated by the pulsing “>” in the upper portion of the
display and Green flashing LED in the top of the door.
Allow Test to run for a minimum of 10 minutes and verify “Ave.:” is
100.00 ml +/- 5%.
1.4.6.1 Alternate Flow Test
Connect patient end of IV Set to a graduated 50 ml cylinder
with 0.1 ml resolution. Set Pump Rate to 100 ml and VTBI to
25 ml, verify Ready to Start is flashing in lower right of the
display. Press the Start/Stop Channel A (Channel B for
SideCar) key and verify the pump begins running as indicated
by the pulsing “>” in the upper portion of the display and
Green flashing LED in the top of the door.
Press the Start/Stop Channel A (Channel B for SideCar) key
when the pump completes the 25 ml and verify graduated
cylinder contains 25 ml +/- 1.25 ml.