15.3.4 Receipt timeout
Signal
Duration
min.
max.
Receive timeout T
1
10 ms
2000 ms
If errors occur in the communication, receiving is terminated after a defined time. The data received to this
point is then discarded. The possible error causes may be:
The cable was unplugged or mechanically damaged during the transfer.
Transmission of the data was prematurely terminated due to a technical fault.
An error occurred in the transfer of the length information, so that the information is incorrectly
transferred. The device then presumes an incorrect overall length of the data.