TIGER XTS Instrument User Manual V1.0
P
ion
eering Gas Sensing Technology.
Page 72 of 74
ionscience.com
Technical Specifications
Minimum Resolution (ppm)
Standard running mode: 0.1 ppm
TAC mode: 0.01 ppm
Tube mode: 0.01 ppm (res. 0.001 ppm)
Minimum Resolution (ppb)
Standard running mode: 0.001 ppm
TAC mode: 0.001 ppm
Tube mode: 0.01 ppm (res. 0.001 ppm)
Maximum Reading
Standard running mode: up to 20,000 ppm or 20,000 mg/m
3
(gas dependent)
Tube mode: 200 ppm or 639 mg/m
3
benzene
Response Time
130 seconds at 20
o
C (variable)
Progressive indication of benzene breakthrough is displayed in
real time
Intrinsically Safe Approvals
•
II 1G Ex ia IIC T4 Ga
•
Tamb = - 15
o
C ≤Ta ≤+45
o
C (with Lithium ion Battery Pack)
•
Tamb = - 15
o
C ≤Ta ≤+4
0
o
C (with Alkaline Battery Pack)
•
IECEx ITS 22.0025X ITS-I22ATEX35111X
•
ITS22UKEX0635X
•
3193491 conforms to UL Std. 913, 61010-1 &
Certified to CAN/CSA Std. C22.2 No. 61010-1
Accuracy
+/- 10 % display r/- one digital benzene
Linearity
+/- 5 % display r/- one digital
Battery Life
Li-ion: Typically up to 24 hours
Typical charge time of 8 hours
Alkaline 3 x AA: Typically 8.5 hours life
Lamps
10.0 eV Krypton PID lamp
Data Logging
>120,000 data log points including date and time stamp
Alarms
Flashing LEDs’s amber (low alarm) red (high alarm)
Sounder 95 dBA at 300 mm (12”)
Vibration on alarm
Pre-programmed TWA and STEL
Flow Rate
≥
220 ml/min