The lower alarm setting should never be greater than the upper alarm. Therefore,
if the lower alarm is set to higher than the upper alarm, the upper alarm is
automatically incremented to the lower alarm level plus 1.
•
Units
Refer to the
topic for details of these options. Select from the
available units by using the
Up
and
Down
keys and then pressing
Enter
.
Zeroing
Pressing the
Zero
soft key presents two Zero options
. Use the
Up
or
Down
keys to
select them. The upper symbol represents an absolute Zero. The lower symbol represents a relative
Zero, which follows the drift of the MiniPID 2 Sensor.
The screen displays a tick '
✓
' to confirm your selection. If the relative Zero has been selected, TIGER
XTS will then Zero itself.
Peak Hold
Press the
Peak Hold
soft key to display the peak reading:
. This reading will stay on
screen, updating as the current peak reading changes. Press the
Reset
soft key to reset the peak
value. Press
Esc
to cancel
.
Information
Pressing the
Information
soft key provides access to a list of screens which display the current
settings of the TIGER XTS. Use the
Up
and
Down
keys to scroll through the list and press
Enter
to
view a screen. Press
Esc
to return to the main running screen.
First screen
Gas selected
Response Factor
Upper alarm
Lower alarm
Second screen
Lamp selected
Date of Factory Calibration
Date of Custom Calibration
Third screen
SPAN 1
SPAN 2
PID live response