3-4
c. Remove oxygen adapter Keeper from rear of monitor using 9/64 inch allen wrench.
d. Unplug Anesthetic Oxygen Sensor Adapter Sub-Assembly from “WASTE GAS
EXHAUST” port on rear panel of monitor.
e. Plug new Anesthetic Oxygen Sensor Adapter Sub-Assembly (or old part if not being
replaced) into the “WASTE GAS EXHAUST” port on the rear panel of the monitor.
f.
Reinstall Oxygen Adapter Keeper to rear of monitor using 9/64 inch allen wrench. (Ensure
Keeper “fingers” engage in and around machined groove in Oxygen Sensor Adapter
housing, and threaded port on Adapter Sub-Assembly is facing upwards.)
g. Screw replacement oxygen sensor into the threaded port on the Adapter Sub-Assembly.
(Hand tighten only.)
h. Plug Adapter Cable Assembly into top of disposable oxygen sensor.
3.8
HE61 Radio Antenna Removal/Replacement.
Remove the optional HE61 radio antenna
as follows:
a. Unscrew Antenna HE61 from Telemetry Antenna Cable AC331.
b. Reverse the above step to install a replacement antenna.
Summary of Contents for Millennia 3155A
Page 2: ......
Page 4: ......
Page 12: ...viii THIS PAGE INTENTIONALLY LEFT BLANK...
Page 32: ...2 12 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 52: ...5 10 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 54: ...6 2...
Page 55: ...6 3...
Page 56: ...6 4...
Page 57: ...6 5...
Page 58: ...6 6...
Page 59: ...6 7...
Page 60: ...6 8...
Page 61: ...6 9...
Page 62: ...6 10...
Page 63: ...A 1 APPENDIX A LIST OF SYMBOLS...
Page 64: ...A 2 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 68: ...C 2 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 71: ...NOTES...
Page 72: ......
Page 73: ......
Page 74: ......
Page 75: ......
Page 76: ......