![Invitrogen E-Gel Power Snap Plus User Manual Download Page 30](http://html1.mh-extra.com/html/invitrogen/e-gel-power-snap-plus/e-gel-power-snap-plus_user-manual_2085413030.webp)
Modify a run
The E
‑
Gel
™
protocol can be cancelled or modified during the run. however the device does not allow
the duration to exceed the maximum allowable run time for the specific E
‑
Gel
™
protocol.
Pause the run
1.
Select
Pause run
to temporarily stop the run.
2.
Select
Resume
to restart the run.
Cancel the run
1.
Select
Pause run
to temporarily stop the run.
2.
Tap the status dial, then select
Cancel run
to stop the run.
Edit gel duration
1.
Select
Pause run
to temporarily stop the run.
2.
Tap the status dial, then select
Edit gel duration
.
3.
Adjust the protocol duration using the
+/–
buttons or tap the duration field to open a keyboard to
enter a number.
4.
Select
Resume
to restart the run.
Note:
Do not run the same gel multiple times or extend the gel protocol beyond the maximum
allowed duration. Running the gel past the allowed duration will damage the gel and result in poor
sample separation.
Change to another protocol
1.
Select
Pause run
to temporarily stop the run.
2.
Tap the status dial, then select
Cancel run
to stop the run.
3.
Select
Set up run.
4.
Select another E
‑
Gel
™
protocol (e.g.,
Reverse E
‑
Gel
™
).
Use the up/down arrows to navigate
through the menu.
5.
Select
Start run
Chapter 2
Methods
Using the E
‑
Gel
™
Power Snap Plus Electrophoresis Device
2